The HearTwave II system is the next generation
Microvolt T-Wave Alternans (MTWA) testing platform simplifying
hospital and office-based SCD risk stratification while
eliminating the need for a host stress system.
Safety and Performance Standards |
| In compliance with |
CAN/CSA 22.2, EN60601-1, EN60601-1-1, EN60601-1-2. Meets the American National Standards for Diagnostic Electrocardiographic Devices (ANSI/AAMI EC-11, EC-12, EC-13, EC-53). |
| Chassis leakage |
< 100 microamps |
| Patient leakage |
< 10 microamps (isolated patient module with Type CF isolation barrier) |
Physical and Environmental Specifications |
| Size |
30" x 52 " x 20 " |
| Shipping weight |
240 lbs |
| Power Requirements |
100-240 V (±10%; 50/60 Hz) |
| Power Consumption |
600 VA (max.) |
| Operating temperature |
+10°C to 32.5°C (50°F to 91°F) |
| Humidity |
30% to 80% non-condensing
|